Radioprotective Effects of a Semicircular X-ray Shielding Device for Operators During CT Fluoroscopy-Guided Interventional Procedures: Experimental and Clinical Studies

CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY(2023)

引用 0|浏览6
暂无评分
摘要
Purpose To evaluate, experimentally and clinically, the radioprotective effects of a semicircular X-ray shielding device for operators during CT fluoroscopy-guided IR procedures. Materials and Methods During experimentation, the reduction rates of scattered radiation rates from CT fluoroscopy were evaluated using a humanoid phantom. Two shielding device positions were tested: “shielding close to the CT gantry” and “shielding close to the operator”. The scattered radiation rate without shielding was also evaluated. The clinical study retrospectively evaluated the operator’s radiation exposure during 314 CT-guided IR procedures. With a semicircular X-ray shielding device (with shielding group, n = 119) or without it (no shielding group, n = 195), CT fluoroscopy-guided IR procedures were performed. Radiation dose measurements were taken using a pocket dosimeter placed near the operator’s eye. For shielding and no shielding groups, the procedure time, dose length product (DLP), and the operator’s radiation exposures were compared. Results Experimentation revealed the respective mean reduction rates of “shielding close to the CT gantry” and “shielding close to the operator” as 84.3% and 93.5% compared with the no-shielding setting. Although no significant differences were found in the procedure time and the DLP between “no shielding” and “with shielding” groups in the clinical study, the operators’ radiation exposure in the “with shielding” group (0.03 ± 0.04 mSv) was significantly lower than in the “no shielding” group (0.14 ± 0.15 mSv; p < .001). Conclusion The semicircular X-ray shielding device provides valuable radioprotective effects for operators during CT fluoroscopy-guided IR.
更多
查看译文
关键词
radioprotective effects,x-ray,fluoroscopy-guided
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要