A GC-FID validated method for detection and quantification of ethylene oxide in urine bags

Jose Filipe Bacalhau Rodrigues, Juliana Alves da Silva, Rebeca Peixoto Medeiros, Joao Vitor Souto de Araujo Queiroz, Maria Roberta de Oliveira Pinto, Solomon Kweku Sagoe Amoah,Marcus Vinicius Lia Fook

MATERIA-RIO DE JANEIRO(2023)

引用 0|浏览9
暂无评分
摘要
The urine bag is a thermo-labile medical-hospital device used in clinical/surgical procedures in urine drainage and collection. As it is a thermolabile material, it is commonly sterilized by ethylene oxide, a highly toxic, mutagenic and carcinogenic cyclic ether gas. Currently, several countries require, through documents and standards, the quantification of ethylene oxide in medical and hospital devices. Thus, this work describes the development and validation of a method employed for the quantification of ethylene oxide (ETO) in urine bags. The method was developed based on the guidelines of ISO 10993-7, ANVISA and INMETRO. The work range, linearity, limits of detection and quantification, repeatability, intermediate precision and selectivity were evaluated. Statistical tools such as Levene's test, Grubbs, residual analysis, F-test of two variances, paired sample T test and ANOVA table, were also applied to evaluate the method. The linearity of the work range (10 - 400 mg.L-1) showed an adequate correlation coefficient (r > 0.9993), with a homoscedastic profile with absence of outliers while the limit of detection and quantification were 1,95 and 6,5 mg.L-1 respectively. After validation, 45 samples of urine bags of different batches were evaluated, which demonstrated levels of ETO below the limit of detection (1,95 mg.L-1). The results highlight a simple method that meets several regulations with a wide working range, high sensitivity and capability to quantify ETO not only in urine bags but also in other medical devices.
更多
查看译文
关键词
ethylene oxide,urine,gc-fid
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要