Model-Based Pose Estimation of Steerable Catheters under Bi-Plane Image Feedback

2023 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, ICRA(2023)

引用 0|浏览7
暂无评分
摘要
Small catheters undergo significant torsional deflections during endovascular interventions. A key challenge in enabling robot control of these catheters is the estimation of their bending planes. This paper considers approaches for estimating these bending planes based on bi-plane image feedback. The proposed approaches attempt to minimize error between either the direct (position-based) or instantaneous (velocity-based) kinematics with the reconstructed kinematics from bi-plane image feedback. A comparison between these methods is carried out on a setup using two cameras in lieu of a bi-plane fluoroscopy setup. The results show that the position-based approach is less susceptible to segmentation noise and works best when the segment is in a non-straight configuration. These results suggest that estimation of the bending planes can be accompanied with errors under 30 degrees. Considering that the torsional buildup of these catheters can be more than 180 degrees, we believe that this method can be used for catheter control with improved safety due to the reduction of this uncertainty.
更多
查看译文
关键词
robotic catheter,bi-plane imaging,pose estimation,catheter navigation