Manipulating Bandwidth of Cryogenically Cooled Receive-Only MRI Detectors

IEEE Sensors Letters(2023)

引用 0|浏览12
暂无评分
摘要
Cryogenically cooled magnetic resonance imaging (MRI) coils demonstrate extremely high Q-factors, leading to a very narrow bandwidth. This bandwidth can be increased in a controlled manner using a preamplifier decoupling. In this letter, an MRI coil design procedure is explained, and an example showing over two times bandwidth increase in a cryogenic coil for 13C isotope imaging in 3-Tesla MRI systems is demonstrated. The achieved wider bandwidth preserves the spectral components of magnetic resonance signals and makes coils less susceptive to detuning due to parasitic coupling to surrounding structures. The presented method is also useful in other applications of high-Q coils, for instance, in low-field MRI systems.
更多
查看译文
关键词
Microwave/millimeter wave sensors,sensor interface circuits,coils,cryogenic electronics,high-Q coils,magnetic resonance imaging (MRI) detectors,noise figure,noise matching,preamplifier decoupling