Performance of micro-beam X-ray absorption spectroscopy at beamline 7.2W of Synchrotron Light Research Institute

Radiation Physics and Chemistry(2023)

引用 0|浏览13
暂无评分
摘要
The existing BL7.2W beamline at SLRI, initially developed for Macromolecular Crystallography, was upgraded with a Si-drift X-ray detector, sample micro-stage, and polycapillary X-ray lens to provide possibilities for experiments using X-ray fluorescence (XRF) imaging and micro-beam X-ray absorption spectroscopy (μXAS) techniques. The micro X-ray beam was characterized. The beam diameter of approximately 25 μm was determined from the knife-edge scan method. The measured photon flux was 3.8 × 109, 1.8 × 109, and 6.4 × 108 photons/s for 8.0, 10.0, and 12.0 keV X-rays, respectively. Imaging XRF and μXAS measurements were successfully performed on a welding joint of oxygen-free high thermal conductivity copper and AISI304 stainless steel. This opens opportunities for SLRI users to utilize this new measurement setup for elemental and chemical characterizations of small areas on various samples.
更多
查看译文
关键词
X-ray absorption spectroscopy,Micro XAS,Synchrotron X-ray,OFHC copper,304 stainless steel,Welding