Utility of Intraoperative Radiographs in Pediatric Cochlear Implant Surgery.

Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery(2023)

引用 0|浏览1
暂无评分
摘要
The use of Cochlear's Slim Modiolar electrode was associated with a significantly increased risk of abnormal intraoperative X-ray compared to the CA electrode. Given the risk of fold-over with routine insertion and normal electrical testing using the Slim Modiolar electrode, we recommend routine use of intraoperative skull X-ray to confirm electrode position.
更多
查看译文
关键词
X-ray,neural response telemetry,pediatric cochlear implant,tip fold-over