Low molecular weight (LWM) hyaluronic acid (HA) and chondroitin sulfate (CS) have a wide range of applications. To determine their molecular weight (MW), we developed a gel permeation chromatography (GPC) method, which is calibrated based on serrated peaks in the chromatograms. MW calibrants were obtained from the enzymolysis of HA and CS using hyaluronidase. The identical structure of calibrants and samples ensured the soundness of the method. The highest confidence MWs were up to 14,454 and 14,605 for HA and CS, respectively, and the standard curves showed very high correlation coefficients. Thanks to the changeless relationship between MW and its contribution to the GPC integral, the second calibration curves could be derived via one GPC column, also embodied correlation coefficients of >0.9999. The discrepancies of MW values were minuscule, and the measurement of a sample could be conducted in <30 min. The accuracy of the method was verified using LWM heparins, and the measured Mw values showed a 1.2 %-2.0 % error relative to pharmacopeia results. The MW results obtained for LWM-HA and LWM-CS samples were also consistent with the results obtained by multiangle laser light scattering. The method was also verified be able to measure the very low MWs.