Development of hard X-ray spectrometer with full digital data acquisition for runaway electron studies at HL-2M

JOURNAL OF INSTRUMENTATION(2023)

引用 1|浏览16
暂无评分
摘要
The development of hard X-ray (HXR) spectrometers with full digital data acquisition for runaway electron studies on the HL-2M tokamak is presented. Consisting of large-sized NaI:Tl detectors and high-performance multichannel analyzers, this system can realize the measurement of HXR with energies ranging from 0.5 to 10 MeV, and spectrum temporal resolution of 1 ms. An average energy resolution of 56 keV at 662 keV has been achieved at counts rates of about 100 kcps. The spectra data is acquired by a digital-based data acquisition system with fast digitization and software signal processing technology. The energy calibrations are performed by using several radioisotope gamma-ray sources, and the detection efficiencies are simulated by MCNP code. First experimental results from HL-2M Ohmic discharges are presented in this paper.
更多
查看译文
关键词
Nuclear instruments and methods for hot plasma diagnostics,Gamma detectors (scin- tillators,CZT,HPGe,HgI etc),X-ray detectors