New uniaxial pressure cell used in a study of magnetization of the Heusler Ni2MnSn-based alloy

J. Kamarad,J. Kastil, M. Misek, Z. Arnold

HIGH PRESSURE RESEARCH(2023)

引用 0|浏览4
暂无评分
摘要
A versatile type of non-magnetic uniaxial pressure cell has been designed for magnetic measurements in SQUID magnetometer. An advanced technology of machining of miniature parts of the cell was used to make new type of an internal squeezer module, where a sample can be compressed by force up to 3 kN in direction perpendicular to external magnetic field. A correct operation of the cell was tested by study of magnetic behaviour of the Heusler off-stoichiometric Ni2MnSn alloy under uniaxial compression. An unexpected significant difference between reversible effects of uniaxial compression on magnetization of the austenite and martensite phases of the alloy is discussed. The uniaxial compression effects are compared with the effects of hydrostatic pressure on a coupling between magnetization and structural transformation of the alloy.
更多
查看译文
关键词
Pressure cell,uniaxial compression,magnetization,Heusler alloy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要