Coupling in-syringe kapok fiber-supported liquid-phase microextraction with flow injection-mass spectrometry for rapid and green biofluid analysis: Determination of antidepressants as an example

JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS(2023)

引用 1|浏览0
暂无评分
摘要
Quantification of substances in biofluid samples (e.g., urine, blood, and cerebrospinal fluids) are useful for clinical diagnosis. In current study, a rapid and green strategy by coupling in-syringe kapok fiber-supported liquid-phase microextraction with flow-injection mass spectrometry was proposed. The natural kapok fiber was used as an oily extraction solvent (e.g., n-octanol) support material, and an in-syringe extraction device was conveniently constructed. The whole extraction processes, including sampling, washing and desorption, were conveniently conducted by simply pulling/pushing the syringe plunger, enabling rapid analyte enrichment and sample purification. The follow-up flow injection-mass spectrometry detection enabled rapid and high throughput analysis. As an example, the proposed method was applied to analyze antidepressants in plasma/ urine, showing satisfied linearities (R2 >= 0.993) in ranges of 0.2-1000 ng/mL. By employing the in-syringe extraction method prior to flow injection-mass spectrometry detection, the LOQs in plasma and urine were reduced by 25-80 folds and 5-25 folds, respectively. Besides, by employing ethanol and 80% ethanol as the desorption solvent and carrier solvent, respectively, the analytical method showed excellent greenness. In gen-eral, the integrated method provides a promising choice for rapid and green biofluid analysis.
更多
查看译文
关键词
green biofluid analysis,antidepressants,in-syringe,fiber-supported,liquid-phase,injection-mass
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要