Five-year polyethylene cup migration and PE wear of the Anatomic Dual Mobility acetabular construct

Archives of Orthopaedic and Trauma Surgery(2023)

引用 2|浏览5
暂无评分
摘要
Introduction Dual mobility implants have been successful in reducing postoperative hip dislocation but mid-term results of cup migration and polyethylene wear are missing in the literature. Therefore, we measured migration and wear at 5-year follow-up using radiostereometric analysis (RSA). Materials and methods A cohort of 44 patients (mean age 73, 36 female) with heterogeneous indications for hip arthroplasty but all with a high risk of hip dislocation received total hip replacement (THA) with The Anatomic Dual Mobility X3 monoblock acetabular construct and a highly crosslinked polyethylene liner. RSA images and Oxford Hip Scores were obtained perioperatively and 1, 2, and 5 years postoperatively. Cup migration and polyethylene wear were calculated using RSA. Results Mean 2-year proximal cup translation was 0.26 mm (95% CI 0.17; 0.36). Proximal cup translation was stable from 1- to 5-year follow-up. Mean 2-year cup inclination (z-rotation) was 0.23° (95% CI − 0.22; 0.68) and was greater in patients with osteoporosis compared to patients without osteoporosis ( p = 0.04). Using 1-year follow-up as baseline, the 3D polyethylene wear rate was 0.07 mm/year (0.05; 0.10). Oxford hip scores improved 19 (95% CI 14; 24) points from mean 21 (range 4; 39) at baseline, to 40 (9; 48) 2 years postoperatively. There were no progressive radiolucent lines > 1 mm. There was 1 revision for offset correction. Conclusions Anatomic Dual Mobility monoblock cups were well-fixed, the polyethylene wear rate was low, and the clinical outcomes were good until 5-year follow-up suggesting good implant survival in patients of different age groups and with heterogeneous indications for THA.
更多
查看译文
关键词
Dual mobility, Hip arthroplasty, HXLPE, Polyethylene wear, Cup fixation, Radiostereometric analysis, Wear rate
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要