Transcutaneous carbon dioxide monitoring during prolonged apnoea with high-flow nasal oxygen.

Acta anaesthesiologica Scandinavica(2023)

引用 3|浏览14
暂无评分
摘要
Transcutaneous carbon dioxide monitoring provided a clinically acceptable substitute for arterial blood gases but as hypercapnia developed to considerable levels, we observed overestimation at high carbon dioxide tensions in patients undergoing apnoeic oxygenation with high-flow nasal oxygen.
更多
查看译文
关键词
apnoeic oxygenation,carbon dioxide,high-flow nasal oxygen,transcutaneous monitoring