Radiotherapy-induced malfunctions of cardiac implantable electronic devices: A meta-analysis.

Heart rhythm(2023)

引用 1|浏览18
暂无评分
摘要
Our meta-analysis suggests that ICDs are more likely to be affected by RT than PMs. These adverse events, especially reset, in patients with cancer were associated with neutron-generating RT and beam energy ≥10 MV. Given the increasing requirement for RT in several patients with cancer as well as the increasing implantation rates of CIEDs, a better risk stratification is needed in this setting.
更多
查看译文
关键词
Cardiac implantable electronic devices,Implantable cardioverter-defibrillators,Malfunction,Meta-analysis,Pacemakers,Radiation therapy