Activity Measurement of 44 Sc and Calibration of Activity Measurement Instruments on Production Sites and Clinics.

Molecules (Basel, Switzerland)(2023)

引用 2|浏览12
暂无评分
摘要
Sc is a promising radionuclide for positron emission tomography (PET) in nuclear medicine. As a part of the implementation of a production site for Sc, precise knowledge of the activity of the product is necessary. At the Paul Scherrer Institute (PSI) and the University of Bern (UniBE), Sc is produced by enriched CaO-target irradiation with a cyclotron. The two sites use different techniques for activity measurement, namely a dose calibrator at the PSI and a gamma-ray spectrometry system at UniBE and PSI. In this work, the Sc was produced at the PSI, and samples of the product were prepared in dedicated containers for onsite measurements at PSI, UniBE, and the Institute of Radiation Physics (IRA) in Lausanne for precise activity measurement using primary techniques and for the calibration of the reference ionization chambers. An accuracy of 1% was obtained for the activity measurement, allowing for a precise calibration of the dose calibrator and gamma-ray spectrometry of the two production sites. Each production site now has the capability of measuring Sc activity with an accuracy of 2%.
更多
查看译文
关键词
Sc-44 production,dose calibrator,medical cyclotron,positron emission tomography,precise activity measurement,precise instruments calibration,γ-ray spectrometry