Characterization of LiF:Mg,Ti thermoluminescence detectors in low-LET proton beams at ultra-high dose rates.

Physics in medicine and biology(2023)

引用 5|浏览10
暂无评分
摘要
This work aims at characterizing LiF:Mg,Ti thermoluminescence detectors (TLDs) for dosimetry of a 250 MeV proton beam delivered at ultra-high dose rates (UHDR). Possible dose rate effects in LiF:Mg,Ti, as well as its usability for dosimetry of narrow proton beams are investigated.LiF:Mg,Ti (TLD-100Microcubes, 1 mm × 1 mm × 1 mm) was packaged in matrices of 5 × 5 detectors. The center of each matrix was irradiated with single-spot low-LET (energy >244 MeV) proton beam in the (1-4500) Gy saverage dose rates range. A beam reconstruction procedure was applied to the detectors irradiated at the highest dose rate (Gaussian beam sigma <2 mm) to correct for volumetric averaging effects. Reference dosimetry was carried out with a diamond detector and radiochromic films. The delivered number of protons was measured by a Faraday cup, which was employed to normalize the detector responses.The lateral beam spread obtained from the beam reconstruction agreed with the one derived from the radiochromic film measurements. No dose rates effects were observed in LiF:Mg,Ti for the investigated dose rates within 3% (= 1). On average, the dose response of the TLDs agreed with the reference detectors within their uncertainties. The largest deviation (-5%) was measured at 4500 Gy s.The dose rate independence of LiF:Mg,Ti TLDs makes them suitable for dosimetry of UHDR proton beams. Additionally, the combination of a matrix of TLDs and the beam reconstruction can be applied to determine the beam profile of narrow proton beams.
更多
查看译文
关键词
FLASH dosimetry,LiF:Mg,Ti,dose rate dependence,proton dosimetry,thermoluminescence detectors
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要