The Integration of Reference Electrode for ISFET Ion Sensors Using Fluorothiophenol-Treated rGO.

Biosensors(2023)

引用 1|浏览15
暂无评分
摘要
Ion-sensitive field-effect transistors (ISFETs) detect specific ions in solutions that enable straightforward, fast, and inexpensive sensors compared to other benchtop equipment. However, a conventional reference electrode (RE) such as Ag/AgCl is limited on the miniaturization of the sensor. We introduce reduced graphene oxide (rGO), which serves as a new RE, when fluorinated (F-rGO) using fluorothiophenol through the π-π interaction. The circular RE is integrated between a fabricated microscale two-channel ISFET, which is capable of detecting two kinds of ions on an indium tin oxide (ITO) thin-film substrate, using the photolithography process. F-rGO bound to this circular region to function as an RE in the ISFETs sensor, which operated stably in solution and showed a relatively high transconductance (g) value (1.27 mS), low drift characteristic (3.2 mV), and low hysteresis voltage (±0.05 mV). It detected proton (H) ions in a buffer solution with high sensitivity (67.1 mV/pH). We successfully detected Na (62.1 mV/dec) and K (57.6 mV/dec) ions in human patient urine using a two-channel ISFET with the F-rGO RE. The F-rGO RE will be a suitable component in the fabrication of low-cost, mass-produced, and disposable ISFETs sensors.
更多
查看译文
关键词
fluorinated graphene oxide,fluorothiophenol,indium tin oxide,ion sensor,ion-sensitive field-effect transistors,reference electrode