MRI conditional safety of mixed systems: Study of the impact of active implant medical device impedance on the radiofrequency-induced lead heating.

Abdou Khadir Fall, Anda Raluca Guraliuc,Nicolas Monfret, Romano Setzu

Magnetic resonance in medicine(2023)

引用 0|浏览1
暂无评分
摘要
The results of our experimental analysis show that different pulse generators would not change the RF-induced heating behavior of a given lead, provided that the pulse generator impedances are small (in our study less than 5%) compared to the overall system impedance.
更多
查看译文
关键词
MRI,RF-induced lead heating,active implant medical device,cardiac implantable electronic device,implantable pulse generator,tier 3
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper