40 MHz Swept-Source Optical Coherence Tomography at 1060 nm Using a Time-Stretch and Supercontinuum Spectral Broadening Dynamics

IEEE Photonics Journal(2022)

引用 0|浏览8
暂无评分
摘要
We present an akinetically swept light source based on a spectrally broadened mode-locked laser in all normal dispersion optical fiber followed by a linear dispersive time stretch and a regenerative amplification stage. Developed for ultrafast swept-source optical coherence tomography (SS-OCT), the source enables the demonstration of A-scan rate at 40 MHz with a 12 μm axial resolution. Through A-scans captured from a fast-moving object, we demonstrate that high-speed OCT is possible at 1 μm wavelength in an architecture which is not intrinsically bandwidth limited. The source therefore shows a promising route to achieving high axial resolution in high-speed swept-source OCT.
更多
查看译文
关键词
OCT,supercontinuum generation,fast imaging,swept source