COMPARISON OF THE 64- AND 80-DETECTOR ROW COMPUTED TOMOGRAPHY AMONG THE CT NUMBER AND RADIATION DOSE DURING LOWER EXTREMITY COMPUTED TOMOGRAPHY ANGIOGRAPHY.
Radiation protection dosimetry(2023)
摘要
To compare the computed tomography (CT) number and the radiation dose between the 64 (group A) and 80-detector row (group B) during lower extremity computed tomography angiography (LE-CTA). We enrolled 144 patients underwent LE-CTA and compared the CT number for the popliteal arteries, radiation dose and the rate of the optimal CT number during the LE-CTA exceeding 200 HU between the two groups. The CT number for the popliteal arteries and mean dose-length product was significantly higher in Group A than in Group B (P < 0.01). The rate of the optimal CT number for the popliteal arteries was 23.6% with Group B scanner and 56.9% with Group A (P < 0.05). The 64-detector row CT was significantly higher in the CT number for the popliteal arteries, radiation dose and rate of the optimal CT number during the LE-CTA than the 80-detector row. Depiction ability did not improve by using a high CT scanner with a wider detector during LE-CTA.
更多查看译文
关键词
tomography angiography,80-detector number,radiation dose