Craniospinal Irradiation: A Dosimetric Comparison Between O-Ring Linac and Conventional C-arm Linac

Advances in Radiation Oncology(2023)

引用 1|浏览4
暂无评分
摘要
Purpose The aim of this study was to perform a dosimetric evaluation between craniospinal irradiation volumetric modulated arc therapy plans designed for an O-Ring and a conventional C-arm Linac. Methods and Materials Two adult patients were selected for this study. Two plans were designed one for a TrueBeam Edge and one for Halcyon O-ring Linac for each patient. The evaluation of the plans was conducted in terms of dose volume histogram analysis of the target volume and organs at risk (OARs) along with total plan monitor units and beam-on time. Paired sample t test was performed to compare Dmax and Dmean of OARs for each plan's comparison. The delivery of volumetric modulated arc therapy plans was evaluated using Octavius 4D phantom. Results All plans demonstrated dose distributions with sufficient planned target volume conformity and homogeneity. The Homogeneity Index and Conformity Index for all plans were found comparable. The paired sample t test did not demonstrate significant difference in terms of Dmax and Dmean of OARs between plans for both patients. All plans met the threshold of 90%, with Halcyon plans having higher gamma passing rates. Conclusions Craniospinal irradiation plans generated for Halcyon and Edge are equivalent in terms of plan quality and dose sparing at OARs. The small variations may have originated from the differences in beam profile or mean energy, the degree of the modulation for each plan and characteristics of multi leaf collimators for each treatment unit. Halcyon is able to deliver a distinctly faster treatment, but Edge provides an automatic rotational correction for patient positioning.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要