Feasibility and Accuracy of Measuring Carotid Plaque Volume (Burden) With Contrast-Enhanced Tomographic 3D Ultrasound and Ultrasound Image Fusion.

S K Rogers, A Phair, J Carriera,J Ghosh, F L Bowling, C McCollum

Annals of vascular surgery(2022)

引用 1|浏览3
暂无评分
摘要
BACKGROUND:Stenosis severity has been the indication for carotid endarterectomy (CEA) for 4 decades, but the annual stroke risk in asymptomatic carotid stenosis >70% is under 2%. Atherosclerotic volume has emerged as a risk factor for future stroke, but needs to be measured noninvasively. Tomographic ultrasound (tUS) is a novel technology that assembles 3D images in seconds. We evaluated accuracy of measuring Carotid Plaque Volume (CPV) with tUS in patients undergoing CEA. METHOD:Consecutive patients were imaged immediately before CEA by tUS and contrast-enhanced tUS (CEtUS). CPV was measured using tUS, CEtUS, and a fused images incorporating both tUS and CEtUS by trained vascular scientists. Precise volume of the endarterectomy specimen was measured using Archimedes technique. RESULTS:Mean ± sd (range) CPV in 129 endarterectomy specimens was 0.75 ± 0.43 cm3 (0.10-2.47 cm3). Mean ± sd CPV measured by tUS (n = 114) was 0.87 ± 0.51 cm3, CEtUS (n = 104) was 0.75 ± 0.45 cm3 and with fusion (n = 95) was 0.83 ± 0.49 cm3. Differences between specimen volume and CPV measured by tUS (0.13 ± 0.24 cm3), CEtUS (-0.01 ± 0.21 cm3) or fusion (-0.08 ± 0.20) were clinically insignificant. Intra-/interobserver differences were minimal. CONCLUSIONS:tUS accurately measures CPV with excellent intra-/interobserver agreement. CEtUS improves accuracy if precise CPV measurement is needed for research but tUS alone would be sufficient for population screening.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要