An Oscillator Driven, Time-Resolved Optical Pump/NIR Supercontinuum Probe Spectrometer

Optics Letters(2022)

引用 1|浏览1
暂无评分
摘要
We present a novel, to the best of knowledge, time-resolved, optical pump/NIR supercontinuum probe spectrometer suitable for oscillators. A NIR supercontinuum probe spectrum (850-1250 nm) is generated in a photonic crystal fiber, dispersed across a digital micromirror device (DMD), and then raster scanned into a single element detector at a 5 Hz rate. Dual modulation of pump and probe beams at disparate frequencies permits simultaneous measurement of both the bare reflectance R and its photoinduced change ΔR through lock-in detection, allowing for continuously self-normalized measurement of ΔR/R. Example data are presented on a germanium wafer sample that demonstrate for signals of order ΔR/R ∼ 10, a 2.87 nm spectral resolution and ≲400 fs temporal resolution pre-recompression, and comparable sensitivity to standard time-resolved, amplifier-based pump-probe techniques.
更多
查看译文
关键词
optical pump/nir,oscillator-driven,time-resolved