An extended bore length solid-state digital-BGO PET/CT system: design, preliminary experience, and performance characteristics

John A. Kennedy, Tala Palchan-Hazan, Quentin Maronnier,Olivier Caselles,Frédéric Courbon, Moshe Levy,Zohar Keidar

EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING(2023)

引用 0|浏览6
暂无评分
摘要
Purpose A solid-state PET/CT system uses bismuth germanium oxide (BGO) scintillating crystals coupled to silicon photomultipliers over an extended 32 cm axial field-of-view (FOV) to provide high spatial resolution and very high sensitivity. Performance characteristics were determined for this digital-BGO system, including NEMA and EARL standards. Methods Spatial resolution, scatter fraction (SF), noise equivalent count rate (NECR), sensitivity, count rate accuracy, and image quality (IQ) were evaluated for the digital-BGO system as per NEMA NU 2-2018, at 2 sites of first clinical install. System energy resolution was measured. Bayesian penalized-likelihood reconstruction (BPL) was used for IQ. EARL Standards 2 studies were reconstructed by BPL combined with a contrast-enhancing deep learning algorithm. An Esser PET phantom was evaluated. Three patient examples were obtained with low-dose radiotracer activity: 2 MBq/kg of [ 18 F]FDG ([ 18 F]-2-fluoro-2-deoxy- d -glucose), 2.3 MBq/kg [ 68 Ga]Ga-DOTA-TATE ([dodecane tetra-acetic acid,Tyr 3 ]-octreotate), and 14.5 MBq/kg [ 82 Rb]RbCl ([ 82 Rb]-rubidium-chloride). Total scan times were ≤ 8 min. Results NEMA sensitivity was 47.6 cps/kBq at the axial center. Spatial resolution at 1 cm from the center axis was ≤4.5 mm (filtered back projection) and ≤3.8 mm (ordered subset expectation maximization). SF was 35.6%, count rate accuracy was 2.16%, and peak NECR was 485.2 kcps at 16.9 kBq/mL. Contrast for IQ was 61.1 to 90.7% (smallest to largest sphere) with background variations from 7.6 to 2.1%, and a “lung” error of 4.7%. The average detector energy resolution was 9.67%. Image quality for patient scans was good. EARL Standards 2 criteria were robustly met and Esser phantom features ≥4.8 mm were resolved at 2 min per bed position. Conclusion A solid-state 32 cm axial FOV digital-BGO PET/CT system provides good spatial and energy resolution, high count rates, and superior NEMA sensitivity in its class, enabling fast clinical acquisitions with low-dose radiotracer activity.
更多
查看译文
关键词
PET/CT,Long axial field-of-view,Bismuth germanium oxide,BGO,Solid-state
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要