Co-axial acoustic-based optical coherence vibrometry probe for the quantification of resonance frequency modes in ocular tissue

Scientific reports(2022)

引用 1|浏览3
暂无评分
摘要
We present a co-axial acoustic-based optical coherence vibrometry probe (CoA-OCV) for vibro-acoustic resonance quantification in biological tissues. Sample vibrations were stimulated via a loudspeaker, and pre-compensation was used to calibrate the acoustic spectrum. Sample vibrations were measured via phase-sensitive swept-source optical coherence tomography (OCT). Resonance frequencies of corneal phantoms were measured at varying intraocular pressures (IOP), and dependencies on Young´s Modulus (E), phantom thickness and IOP were observed. Cycling IOP revealed hysteresis. For E = 0.3 MPa, resonance frequencies increased with IOP at a rate of 3.9, 3.7 and 3.5 Hz/mmHg for varied thicknesses and 1.7, 2.5 and 2.8 Hz/mmHg for E = 0.16 MPa. Resonance frequencies increased with thickness at a rate of 0.25 Hz/µm for E = 0.3 MPa, and 0.40 Hz/µm for E = 0.16 MPa. E showed the most predominant impact in the shift of the resonance frequencies. Full width at half maximum (FWHM) of the resonance modes increased with increasing thickness and decreased with increasing E. Only thickness and E contributed to the variance of FWHM. In rabbit corneas, resonance frequencies of 360–460 Hz were observed. The results of the current study demonstrate the feasibility of CoA-OCV for use in future OCT-V studies.
更多
查看译文
关键词
Biomedical engineering,Biophotonics,Eye diseases,Imaging and sensing,Science,Humanities and Social Sciences,multidisciplinary
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper