Comparison between Fourth-Generation FloTrac/Vigileo System and Continuous Thermodilution Technique for Cardiac Output Estimation after Time Adjustment during Off-Pump Coronary Artery Bypass Graft Surgery: A Retrospective Cohort Study.

Journal of clinical medicine(2022)

引用 2|浏览10
暂无评分
摘要
(1) Background: Previous studies reported limited performance of arterial pressure waveform-based cardiac output (CO) estimation (FloTrac/Vigileo system; CO-FloTrac) compared with the intermittent thermodilution technique (CO). However, errors due to bolus maneuver and intermittent measurements of CO could limit its use as a reference. The continuous thermodilution technique (CO) may relieve such limitations. (2) Methods: The performance of CO-FloTrac was retrospectively assessed using continuous recordings of intraoperative physiological data acquired from patients who underwent off-pump coronary artery bypass graft (OPCAB) surgery with CO monitoring using both CO-FloTrac and CO. Optimal time adjustments between the two measurements were determined based on R-squared values. (3) Results: A total of 134.2 h of data from 30 patients was included in the final analysis. The mean bias was -0.94 (95% CI, -1.35 to -0.52) L/min and the limits of agreements were -3.64 (95% CI, -4.44 to -3.08) L/min and 1.77 (95% CI, 1.21 to 2.57) L/min. The percentage error was 66.1% (95% CI, 52.4 to 85.8%). Depending on the time scale and the size of the exclusion zone, concordance rates ranged from 61.0% to 75.0%. (4) Conclusion: Despite the time adjustments, CO-FloTrac showed non-negligible overestimation, clinically unacceptable precision, and poor trending ability during OPCAB surgery.
更多
查看译文
关键词
arterial pressure waveform,cardiac output,coronary artery bypass graft,hemodynamic monitoring,thermodilution
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要