Paper-based colorimetric glucose sensor using Prussian blue nanoparticles as mimic peroxidase

BIOSENSORS & BIOELECTRONICS(2023)

引用 8|浏览22
暂无评分
摘要
A novel paper-based colorimetric glucose sensor was proposed employing Prussian blue nanoparticles (PB NPs) as mimic peroxidase. The sensor was manufactured by spraying solution containing PB NPs, glucose oxidase and chromogenic agents into a paper, then coating the filter layer and spreading layer on the top. The layer-by-layer structure enabled the sensor detect glucose in whole blood, as well as elimination of the coffee-ring effect which ensure the performance. As a powerful alternative to natural peroxidase, PB NPs showed the mimic enzymatic activity well preserved in dry environment. The manufacture process of the sensor is easy to be industrialized. Under optimal conditions, the sensor exhibited a linear range from 2.5 mM to 25 mM for glucose in blood with satisfactory reproducibility (the coefficient of variant <4%), great storage stability (1 month at 45 C) and excellent linearity compared with those commercial kits (R > 0.99). Coupled with a handhold device, the PB NPs-based test strip realized the goal of personal operation, user-friendly control, automatic readouts, and data storage, and able to link the Cloud, showing unique potential in clinical application, especially in community -level medical scenarios.
更多
查看译文
关键词
Glucose,Paper-based sensor,Prussian blue,Handheld system,Whole blood