Development of a Compact and Portable High-sensitivity Omnidirectional Compton Camera with Detector Rotation Function

2021 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)(2021)

引用 0|浏览4
暂无评分
摘要
We report the development of a compact and portable high-sensitivity omnidirectional Compton camera. This system can visualise radioactive Cesium even when surface or air dose rates is less than 1 μSv/h. This detector is designed for the Fukushima Daiichi Nuclear Power Plant accident, and low-level radioactive contamination in medical facilities. We adopted the Compton camera technology reported in a previous study (Muraishi et al., Jap. J. Appl. Phys. 2020). This technology enables shift-invariant gamma-ray imaging. The detector can be operated by a PC via Wi-Fi with a power consumption of 15 W. Furthermore, by adopting metal package photomultiplier tubes (PMTs) and optimising the crystal arrangement, we made the detector smaller and improved image quality.
更多
查看译文
关键词
portable high-sensitivity omnidirectional Compton camera,Fukushima Daiichi Nuclear Power Plant accident,low-level radioactive contamination,Compton camera technology,detector rotation function,compact sensitivity omnidirectional Compton camera,Wi-Fi,metal package photomultiplier tubes,crystal arrangement