Multi-arrays of 3D cylindrical microdetectors for beam characterization and microdosimetry in proton therapy

Frontiers in Physics(2022)

引用 0|浏览1
暂无评分
摘要
The present work shows the performance of two new large microdosimetry multi-array systems having two different configurations, namely, pixel and strip configurations. They cover radiation sensitive areas of 1.9 cm × 0.1 cm and 5.1 cm × 0.1 cm, respectively. The microdosimetry systems are based on arrays of 3D cylindrical silicon microdetectors. The 3D electrodes are etched inside the silicon and have a 25 μm diameter and a 20 μm depth. Each of these unit cells is completely isolated from the others and has a well defined 3D micrometric radiation sensitive volume. The pixel-type device consists of 25 × 5 independent silicon-based detectors (500 in total), each one connected to a readout channel, collecting information in 2D in the transverse planes to the particle beam direction. The distance between the individual detectors (pitch) is 200 μm in the horizontal axis and 250 μm in the vertical one. In the case of the strip-type system, we have 512 “columns” (or strips) of 10 detectors per column. Each strip is connected to a readout channel, giving us information in one dimension, but with better statistics than a single pixel. In this system, both the horizontal and vertical pitches are 100 μm.Both systems have been tested under proton beam irradiations at different energies between 6 and 24 MeV to obtain the corresponding microdosimetry quantities along the Bragg peak and distal edge. The measurements were performed at the Accélérateur Linéaire et Tandem à Orsay (ALTO, France). The microdosimetry quantities were successfully obtained with spatial resolutions of 100–250 μm. Experimental results were compared to Monte Carlo simulations and an overall good agreement was found. Both microdetector systems showed a good microdosimetry performance under clinical-equivalent fluence rates along distances of several centimeters. This work demonstrates that the two new systems having different configurations can be clinically used as microdosimeters for measuring the lineal energy distributions in the context of proton therapy treatments. Additionally, they could be also used for beam monitoring.
更多
查看译文
关键词
microdosimetry,proton therapy,silicon radiation detectors,beam monitoring,Monte Carlo simulations
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要