Central blood pressure estimation in type 1 diabetes: impact and implications of peripheral calibration method.

Journal of Hypertension(2022)

引用 0|浏览0
暂无评分
摘要
In patients with T1DM, there are significant differences in central BP values estimated with radial artery tonometry, depending on the method used for calibration of the radial waveforms. Brachial MAP/DBP-calibration resulted in consistently higher central BP as compared to using brachial SBP/DBP, leading to patient re-stratification. Hence, the accuracy of noninvasive estimation of central BP by radial tonometry is dependent on calibration approach, and this problem must be resolved in validation studies using an invasive reference standard to determine which method best estimates true central BP.
更多
查看译文
关键词
arterial stiffness,calibration,central blood pressure,tonometry,type 1 diabetes