[Usefulness of an Ultrasound System with Automatic Bladder Urine Volume Measurement Using Artificial Intelligence Technology in Radiotherapy].

Nihon Hoshasen Gijutsu Gakkai zasshi(2022)

引用 1|浏览1
暂无评分
摘要
PURPOSE:We aimed to investigate the usefulness of iViz air ver.4 Convex (FUJIFILM, Tokyo) as a tool to determine the bladder capacity before prostate radiotherapy by comparing it with the existing BladderScan BVI 6100 (Verathon Inc., Bothell, Washington). METHODS:We investigated the usefulness of iViz air as a tool to determine the bladder capacity before prostate radiotherapy by comparing it with the current BladderScan. RESULTS:The absolute value of each error rate was approximately 30.9%±27.2% and 26.4%±18.9% for the BladderScan and iViz air, respectively, with no significant differences between the instruments (p=0.16). Evaluated by urine volume, the mean error rates for bladder volumes >50 ml were 26.9%±19.0% and 26.1%±18.5% for the Bladder Scan and iViz air, respectively, with no significant differences (p=0.56). However, the BladderScan and iViz air had significantly higher error rates of 89.5%±52.5% and 31.5%±25.1%, respectively, if the bladder volume was <50 ml (p=0.005). CONCLUSION:The iViz air has limited measurement error to confirm images, especially in limited volumes, suggesting that it is a useful bladder capacity measurement device in performing prostate radiotherapy.
更多
查看译文
关键词
artificial intelligence,bladder urine volume,prostate,radiotherapy,ultrasound