A Conductive Hydrogel Microneedle‐Based Assay Integrating PEDOT:PSS and Ag‐Pt Nanoparticles for Real‐time, Enzyme‐less, and Electrochemical Sensing of Glucose

Advanced Healthcare Materials(2022)

引用 6|浏览2
暂无评分
摘要
Continuous glucose meters (CGMs) have tremendously boosted diabetes care by emancipating millions of diabetic patients' need for repeated self-testing by pricking their fingers every few hours. However, CGMs still suffer from major deficiencies regarding accuracy, precision, and stability. This is mainly due to their dependency on an enzymatic detection mechanism. Here a low-cost hydrogel microneedle (HMN)-CGM assay fabricated using swellable dopamine (DA)-hyaluronic acid (HA) hydrogel for glucose interrogation in dermal interstitial fluid (ISF) is introduced. Platinum and silver nanoparticles are synthesized within the 3D porous hydrogel scaffolds for nonenzymatic electrochemical sensing of the glucose. Incorporation of a highly water dispersible conductive polymer, poly(3,4-ethylenedioxythiophene) polystyrene sulfonate (PEDOT:PSS) enhances the electrical properties of HMN array, making the patch suitable as the working electrode of the sensor. The in vitro and ex vivo characterization of this newly developed HMN patch is fully studied. The performance of the HMN-CGM for real-time measurement of glucose is also shown using a rat model of type 1 diabetes. The device introduces the first HMN-based assay for tracking important disease biomarkers and expect to pave the way for next generation of polymeric-based sensors.
更多
查看译文
关键词
conductive hydrogel,enzyme-less glucose detection,hydrogel microneedles,in situ sensing,platinum and silver nanoparticles
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要