A case of insulinoma-induced hypoglycemia managed by Dexcom G4 Platinum

NEUROENDOCRINOLOGY LETTERS(2022)

引用 0|浏览0
暂无评分
摘要
This report details the case of a 41-year-old woman who was diagnosed with insulinoma. As the patient developed severe life-threatening hypoglycemia, we introduced Dexcom G4 Platinum (DG4P), a modern continuous glucose-monitoring system (CGM). The algorithm of the sensor glucose (SG) values of CGM is based on patients with diabetes; therefore, we evaluated the accuracy of DG4P in this patient. The mean absolute relative differences and absolute differences between SG of DG4P and self-monitoring of blood sugar values were 10.8%+/- 8.3% and 6.8 +/- 5.7 mg/dL, respectively, in the hypoglycemic region, which verifies DG4P's accuracy. DG4P was found to be useful for monitoring hypoglycemia not only in patients with diabetes but also in those with insulinoma.
更多
查看译文
关键词
Diabetes, Insulinoma, Dexcom G4 Platinum, Hypoglycemia, Neuroendocrine tumor