Pressure Measurement in a Bladder Phantom Using Contrast-Enhanced Ultrasonography-A Path to a Catheter-Free Voiding Cystometrogram.

Investigative radiology(2023)

引用 3|浏览8
暂无评分
摘要
Microbubble subharmonic emission is an excellent indicator of bladder phantom pressure variation. The strong correlation between SHAPE signal and bladder phantom pressure is indicative of the applicability of this method in measuring bladder pressure during a CMG. Our results suggest that different SHAPE conversion factors may be needed for different events during a CMG (ie, at different time points of a CMG). These findings will help us better protocolize this method for introduction into human subjects and allow us to take the next step toward developing a catheter-free voiding CMG using SHAPE.
更多
查看译文
关键词
contrast-enhanced ultrasound,contrast media,cystometrogram,microbubbles,pressure-flow studies,subharmonic-aided pressure estimation,SHAPE,ultrasonography,urodynamics