Nature-inspired saccadic-like electrical stimulation paradigm promotes sustained retinal ganglion cell responses by spatiotemporally alternating activation of contiguous multi-electrode patterns

Wadood Haq, Sunetra Basavaraju, Achim Speck,Eberhart Zrenner

JOURNAL OF NEURAL ENGINEERING(2022)

引用 2|浏览0
暂无评分
摘要
Objective. Retinal electrical stimulation using multi-electrode arrays (MEAs) aims to restore visual object perception in blind patients. However, the rate and duration of the artificial visual sensations are limited due to the rapid response decay of the stimulated neurons. Hence, we investigated a novel nature-inspired saccadic-like stimulation paradigm (biomimetic) to evoke sustained retinal responses. For implementation, the macroelectrode was replaced by several contiguous microelectrodes and activated non-simultaneously but alternating topologically. Approach. MEAs with hexagonally arranged electrodes were utilized to simulate and record mouse retinal ganglion cells (RGCs). Two shapes were presented electrically using MEAs: a 6e-hexagon (six hexagonally arranged 10 mu m electrodes; 6e-hexagon diameter: 80 mu m) and a double-bar (180 mu m spaced, 320 mu m in length). Electrodes of each shape were activated in three different modes (simultaneous, circular, and biomimetic ('zig-zag')), stimulating at different frequencies (1-20 Hz). Main results. The biomimetic stimulation generated enhanced RGC responses increasing the activity rate by 87.78%. In the spatiotemporal context, the electrical representation of the 6e-hexagon produced sustained and local RGC responses (similar to 130 mu m corresponding to similar to 2.5 degrees of the human visual angle) for up to 90 s at 10 Hz stimulation and resolved the electrically presented double-bar. In contrast, during conventional simultaneous stimulation, the responses were poor and declined within seconds. Similarly, the applicability of the biomimetic mode for retinal implants (7 x 8 pixels) was successfully demonstrated. An object shape impersonating a smile was presented electrically, and the recorded data were used to emulate the implant's performance. The spatiotemporal pixel mapping of the activity produced a complete retinal image of the smile. Significance. The application of electrical stimulation in the biomimetic mode produced locally enhanced RGC responses with significantly reduced fading effects and yielded advanced spatiotemporal performance reflecting the presented electrode shapes in the mapped activity imprint. Therefore, it is likely that the RGC responses persist long enough to evoke visual perception and generate a seamless image, taking advantage of the flicker fusion. Hence, replacing the implant's macroelectrodes with microelectrodes and their activation in a topologically alternating biomimetic fashion may overcome the patient's perceptual image fading, thereby enhancing the spatiotemporal characteristics of artificial vision.
更多
查看译文
关键词
artificial vision, retinal implants, electrical stimulation, retinal degeneration, bionic prosthesis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要