Implantable Cardioverter-defibrillator Magnetic Interference by the iPhone 12: Is It Clinically Significant?

The Journal of innovations in cardiac rhythm management(2022)

引用 2|浏览8
暂无评分
摘要
The iPhone 12 series (Apple, Inc., Cupertino, CA, USA) contains a circular array of magnets around a central charging coil (compatible with "Magsafe" technology). The device was recently reported to have magnetic interference with implantable cardioverter-defibrillators (ICDs). We sought to test the electromagnetic interference of the iPhone 12 in inhibiting life-saving therapies of ICDs in clinical settings. After obtaining written informed consent, an iPhone 12 was placed over the device generators of 17 patients in the ICD clinic. Device interrogation was performed immediately before and after placing the iPhone over the ICD generator to evaluate for any inhibition of device therapies. To emulate a real-world scenario, the iPhone 12 was not placed directly over the skin above the device generator but instead was positioned over the patients' clothes. None of the device interrogations revealed interruption of device therapies due to the iPhone. We concluded that, despite the iPhone having shown in vitro interference of ICD functioning, its effects are not clinically relevant in vivo. Larger studies need to be performed to confirm this finding and guide safety recommendations regarding the use of iPhones containing magnets by patients with implanted ICDs.
更多
查看译文
关键词
ICD,magnetic interference,sudden cardiac arrest