Mechanical Cardiac Support With an Implantable Direct Cardiac Compression Device: Proof of Concept

ANNALS OF THORACIC SURGERY(2022)

引用 1|浏览2
暂无评分
摘要
PURPOSE We examined the hemodynamic effects of a new, implantable, direct cardiac assist device in an ovine heart failure model.DESCRIPTION The device, which encompasses both left and right ventricles, is inserted through the pericardial apex and self-expands to encompass the heart without suturing. The intact pericardium anchors the device in place. The device has 2 concentric chamber layers: an internal chamber layer filled with fluid to conform to the heart and an external chamber layer filled with air that provides external compression and negative pressure to aid relaxation. EVALUATION The device was implanted in 7 sheep with heart failure induced by microsphere embolization. Cardiac performance was assessed for 6 to 8 hours. The cardiac assist device provided cardiac systolic and diastolic assis-tance, as shown by pressure tracings of the left ventricle and aorta, pulmonary artery flow, and DdP/dt. Central venous pressure decreased during cardiac assistance. No anatomic damage was noted postmortem.CONCLUSIONS Systolic and diastolic cardiac assistance can be achieved with this device that compresses and re-laxes in synchrony with the native cardiac cycle.(Ann Thorac Surg 2022;114:1944-50)(c) 2022 The Authors. Published by Elsevier Inc. on behalf of The Society of Thoracic Surgeons. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).