Gordon Coupler with Inductive or Capacitive Iris for Small EPR Resonators for Aqueous Samples

Mett, Richard R.,Hyde, James S.

Applied Magnetic Resonance(2022)

引用 0|浏览6
暂无评分
摘要
The Gordon coupler was introduced for use in EPR experiments at liquid helium temperatures. It provides an evanescent wave incident on the iris of a microwave resonator. Match of power incident on the coupler to the resonator is obtained by variation of the amplitude of an evanescent wave that arises from displacement of a dielectric wedge in a tapered waveguide. Reduced microphonics from helium bubbling was reported. The Gordon coupler was subsequently extended from cavity resonators to loop-gap resonators, initially at helium temperatures but later for aqueous samples. Plastics with low dielectric constants, usually Teflon, were used. Here, we extend the Gordon coupler for application in X-band five-loop–four-gap resonators using fused quartz, sapphire, or rutile dielectrics, noting that the size of the coupler can then be commensurate with dimensions of dielectric loop-gap resonators as well as dielectric tube resonators. Finite-element modeling of electromagnetic fields has been carried out, and use of a capacitive iris that interfaces with the Gordon coupler reduces pulling of the resonant frequency when matching the resonator.
更多
查看译文
关键词
Gordon coupler,capacitive and resonant iris,dielectric resonator,electron paramagnetic resonance,inductive,loop-gap resonator,sapphire,single crystal rutile