A small-scale realistic inter-laboratory accident dosimetry comparison using the TL/OSL from mobile phone components

Radiation Measurements(2022)

引用 4|浏览6
暂无评分
摘要
An inter-laboratory comparison of a small-scale field experiment was conducted in the framework of the Korea retrospective dosimetry network (KREDOS) to verify the reliability and compatibility of thermoluminescence and optically stimulated luminescence techniques for accident dosimetry. Three institutes participated in this comparison, including the Korea Atomic Energy Research Institute, the Korea Institute of Nuclear Safety and the University of Salzburg. Two adult CIRS torso phantoms were used as virtual subjects exposed to a high-activity Ir-192 source, developing both homogeneous and inhomogeneous exposures in a shielded enclosure. The target materials for dose estimation were mobile phone components, such as resistors, display glasses, and chip cards. LiF:Mg,Cu, Si thermoluminescent (TL) detectors were used as reference dosimeters. The mobile phones were attached to the chest, hip, and thighs of the anthropomorphic phantoms. The measured doses ranged from 100 mGy to 2 Gy, depending on the position of the mobile phone on the phantom. The proficiency of the dose assessment was tested at three laboratories. Using the applied protocols, the local dose distribution on the display glasses was unclear, but the distribution around the body was confirmed.
更多
查看译文
关键词
Accident dosimetry,Thermoluminescence,Optically stimulated luminescence,Inter-laboratory comparison,Small-scale field experiment