Neutron flux parameters at pneumatic irradiation tube Pn-1 of Kyoto University Research Reactor (KUR) were measured in order to establish k 0 -INAA method. The measured thermal to epithermal neutron fluxes ratio ( f ), shape factor of epithermal neutron flux ( α ), and thermal ( ϕ th ) and fast ( ϕ f ) neutron fluxes were 29.7 ± 2.0, − 0.015 ± 0.003, and (3.17 ± 0.11) × 10 12 and (4.79 ± 0.16) × 10 11 cm −2 s −1 at 1 MW, respectively. The neutron flux spectrum showed a remarkable stability over the weekly operation cycle as well as at different reactor powers. The analysis results of reference materials [multi-element liquid standard, and Oriental Basma Tobacco Leaves (INCT-OBTL-5)] revealed that the k 0 -INAA procedure established at the KUR can be regarded as a reliable standardization method of multi-elemental INAA; measured concentrations have zeta-score < ± 2. Corrections due to nuclear interference resulting from fast neutron reactions must be considered for samples with relatively high concentrations of interfering elements. MCNP5 code was used to estimate the total efficiencies of gamma-rays, which might be required to calculate correction factor for the true coincidence effects.