Unexpected high early failure rate of the Nexel total elbow arthroplasty

JSES International(2022)

引用 0|浏览0
暂无评分
摘要
Primary TEA with implantation of this implant was associated with an unacceptably high rate of early implant loosening, periprosthetic fracture, and reoperation. We hypothesize that this early unexpected mechanical failure could be explained by both the utilization of a titanium-on-polyethylene bearing surface and a more posterior center of rotation causing premature anterior impingement with flexion leading to failure of the bonding interface, secondary titanium particle shedding, polyethylene wear, and osteolysis.
更多
查看译文
关键词
Level IV,Case Series,Treatment Study