A Reconfigurable Neural Stimulation IC With a High-Resolution Strength Control and In-Situ Neural Recording Function for Cochlear Implant Systems

IEEE Solid-State Circuits Letters(2022)

引用 3|浏览4
暂无评分
摘要
This letter presents a neural stimulation IC for the 32-channel cochlear implant system. A zoom current DAC (I-DAC), whose full range is reconfigured to fit the interval between the comfort (C) and threshold (T) levels, is introduced for high-resolution stimulation strength control to provide high-fidelity sound perception with minimal hardware overhead. The IC also includes an in-situ neural recording function. The effectiveness of auditory nerve stimulation is validated by monitoring electrically evoked compound action potential (ECAP) in-vivo using a rat model. It is found that the action potential amplitude increases as the stimulation current level becomes larger.
更多
查看译文
关键词
Cochlear implant (CI),electrically evoked compound action potential (ECAP),neural recording,reconfigurable neural stimulation,zoom current DAC