Comparison of the Hemocompatibility of an Axial and a Centrifugal Left Ventricular Assist Device in an In Vitro Test Circuit

JOURNAL OF CLINICAL MEDICINE(2022)

引用 2|浏览13
暂无评分
摘要
Background: Hemocompatibility of left ventricular assist devices is essential for preventing adverse events. In this study, we compared the hemocompatibility of an axial-flow (Sputnik) to a centrifugal-flow (HeartMate 3) pump. Methods: Both pumps were integrated into identical in vitro test circuits, each filled with 75 mL heparinized human blood of the same donor. During each experiment (n = 7), the pumps were operated with equal flow for six hours. Blood sampling and analysis were performed on a regular schedule. The analytes were indicators of hemolysis, coagulation activation, platelet count and activation, as well as extracellular vesicles. Results: Sputnik induced higher hemolysis compared to the HeartMate 3 after 360 min. Furthermore, platelet activation was higher for Sputnik after 120 min onward. In the HeartMate 3 circuit, the platelet count was reduced within the first hour. Furthermore, Sputnik triggered a more pronounced increase in extracellular vesicles, a potential trigger for adverse events in left ventricular assist device application. Activation of coagulation showed a time-dependent increase, with no differences between both groups. Conclusions: This experimental study confirms the hypothesis that axial-flow pumps may induce stronger hemolysis compared to centrifugal pumps, coming along with larger amounts of circulating extracellular vesicles and a stronger PLT activation.
更多
查看译文
关键词
left ventricular assist device, in vitro test circuit, Sputnik, HeartMate 3, hemocompatibility, hemolysis, coagulation, platelet activation, extracellular vesicles
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要