Development and Validation of a Method of Liquid Chromatography Coupled with Tandem Mass Spectrometry for Quantification of ST-246 (Tecovirimat) in Human Plasma

MOLECULES(2022)

引用 5|浏览2
暂无评分
摘要
The aim of this work was to develop and validate a sensitive and robust method of liquid chromatography coupled with tandem mass spectrometry to quantitate ST-246 (tecovirimat) in plasma using an internal standard (2-hydroxy-N-{3,5-dioxo-4-azatetracyclo [5.3.2.02.6.08.10]dodec-11-en-4-yl}-5-methylbenzamide). The method was validated in negative multiple reaction monitoring mode following recommendations of the European Medicines Agency for the validation of bioanalytical methods. The calibration curve for the analyte was linear in the 10-2500 ng/mL range with determination coefficient R-2 > 0.99. Intra- and inter-day accuracy and precision for three concentrations of quality control were <15%. Testing of long-term stability of ST-246 (tecovirimat) in plasma showed no degradation at -20 degrees C for at least 3 months. The method was applied to a clinical assay of a new antipoxvirus compound, NIOCH-14. Thus, the proposed method is suitable for therapeutic drug monitoring of ST-246 (tecovirimat) itself and of NIOCH-14 as its metabolic precursor.
更多
查看译文
关键词
MRM, LC-MS, MS, tecovirimat, ST-246, NIOCH-14, plasma