A Composite Velocity Map Imaging Spectrometer for Ions and 1 keV Electrons at the Shanghai Soft X-ray Free-Electron Laser

APPLIED SCIENCES-BASEL(2021)

引用 3|浏览1
暂无评分
摘要
Velocity map imaging (VMI) spectrometry is widely used to measure the momentum distribution of charged particles with the kinetic energy of a few tens of electronVolts. With the progress of femtosecond laser and X-ray free-electron laser, it becomes increasingly important to extend the electron kinetic energy to 1 keV. Here, we report on a recently built composite VMI spectrometer at the Shanghai soft X-ray free-electron laser, which can measure ion images and high-energy electron images simultaneously. In the SIMION simulation, we extended the electron kinetic energy to 1 keV with a resolution <2% while measuring the ions with the kinetic energy of 20 eV. The experimental performance is tested by measuring Ar 2p photoelectron spectra at Shanghai Synchrotron Radiation Facility, and O+ kinetic energy spectrum from dissociative ionization of O-2 by 800 nm femtosecond laser. We reached a resolution of 1.5% at the electron kinetic energy of 500 eV. When the electron arm is set for 100 eV, a resolution of 4% is reached at the ion kinetic energy of 5.6 eV. This composite VMI spectrometer will support the experiment, such as X-ray multi-photon excitation/ionization, Auger electrons emission, attosecond streaking.
更多
查看译文
关键词
velocity map imaging, Synchrotron Radiation, femtosecond laser, free-electron laser
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要