Frequency and mode measurement techniques for megawatt-class gyrotrons

TM-TECHNISCHES MESSEN(2022)

引用 3|浏览24
暂无评分
摘要
State-of-the-art vacuum electron tubes such as gyrotrons, deliver RF output powers up to more than 2 MW at frequencies up to 170 GHz. In terms of the very high power levels, a proper verification of the gyrotron components itself and measurements during gyrotron operation are vital to prevent possible fatal errors. Several basic RF measurement setups, which are used at IHM/KIT, are discussed. Currently, their upper frequency limit is 175 GHz. In terms of future gyrotron operation above 200 GHz, upgrades of the measurement setups for operation up to 260-330 GHz are prepared. The experimental devices discussed herein are a quasi-optical mode generator for the verification of the quasi-optical gyrotron output system, the window measurement test stand to verify the ceramic gyrotron output window and the frequency diagnostic system to measure the operating frequency and thereby the excited mode.
更多
查看译文
关键词
Gyrotron, RF measurement setup, mode generator, high-order modes, RF windows, frequency diagnostic