Development and evaluation of a thallium (I) bromide dosimeter for intracavitary radiotherapy quality assurance

JOURNAL OF INSTRUMENTATION(2022)

引用 2|浏览4
暂无评分
摘要
In intracavitary radiotherapy, incorrect source locations can result in excessive doses to normal tissues. Therefore, it is essential to accurately evaluate the source location. In this study, we investigated a digital line dosimeter based on thallium (I) bromide (TlBr) to improve the existing analogue verification method. Therefore, a polycrystalline T1Br unit cell dosimeter was manufactured, and the measurement performance of iridium-192 (Ir-192) sources was evaluated. We found that the dosimeter's reproducibility satisfied the evaluation criteria of 1.5% with a relative standard deviation of 1.44%. Moreover, the linearity showed excellent results (linear coefficient, R-2 = 0.9999). The distance dependence showed a difference of 0.03 cm at 50 degrees 7o intensity when compared to the inverse square value, whereas the angular dependence showed a large difference when compared to a diode. As the angle increased, the intensity gradually decreased, resulting in a difference of up to 41.9 degrees 70. These results demonstrate the excellent performance of the TlBr dosimeter. However, because of the significant influence of angular dependence, a measurement distance that minimises this error should be applied when manufacturing line dosimeters in the future.
更多
查看译文
关键词
Radiotherapy concepts, Dosimetry concepts and apparatus, Radiation damage to detector materials (solid state)