Preliminary experience of permanent left bundle branch area pacing using stylet-directed pacing lead without delivery sheath.

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(2022)

引用 1|浏览20
暂无评分
摘要
BACKGROUND:Left bundle branch area pacing (LBBAP) aims to capture the cardiac conduction system in area of the left bundle branch. Currently, LBBAP is mainly performed using lumen-less pacing leads (LLLs) with preshaped sheath. However, the data on LBBAP with stylet-driven leads (SDLs) without sheath is limited. OBJECTIVE:This study presents the feasibility, safety, and pacing characteristics of LBBAP using SDLs without the support of sheath. METHODS:A total of 25 patients with bradycardia indications who received LBBAP implantation with an attempt of SDL (FINELINE II 4471 lead, Boston Scientific, MA, US) between August 2020 and April 2021 at Sir Run Run Shaw Hospital were included in this retrospective cohort study. Twenty of them finally were paced with SDL in priority (SDL-LBBAP group). Twenty propensity score matching patients who underwent LBBAP with LLL (Select Secure 3830 lead, Medtronic, MN, US) and 20 right ventricular septal pacing (RVSP) with regular active fixation lead respectively in the same period (the LLL-LBBAP group and RVSP group) were compared using ECG characteristics, pacing parameters and complications during 6-month follow-up. RESULTS:LBBAP was successful with SDL in 23 of 25 patients (92%) and 20 of them were paced with SDL first. In the SDL-LBBAP group, the average age was 70.4 ± 8.2 years, and 55% of patients were male. Paced QRS duration and the stimulus to peak left ventricular activation time (Sti-LVAT) in SDL-LBBAP group were similar with those in LLL-LBBAP group and significantly shorter than those in RVSP group (126.1±14.1 ms vs. 124.8±10.9 ms, p = 1.00; 77.7 ± 11.2 ms vs. 73.5 ± 9.3 ms, P = .75; 126.1 ± 14.1 ms vs. 147.7 ± 22.5 ms, P<.001; 77.7 ± 11.2 ms vs. 97.0 ± 13.2 ms, P<.001). The pacing threshold and R-wave amplitude of SDL-LBBAP group were 0.53 ± 0.18V and 11.53 ± 3.63 mV at baseline respectively, which were comparable with the other two groups. During the 6-month follow-up, the pacing parameters remained stable and no lead-related complications were recorded. CONCLUSION:It is feasible and safe to use stylet-directed pacing lead for permanent LBBAP without a delivery sheath. Similar to LLL, LBBAP using SDL showed stable parameters and narrower paced QRS duration compared with RVSP, which could be an alternative to LLL in LBBAP.
更多
查看译文
关键词
feasibility, left bundle branch area pacing, lumen-less pacing lead, physiologic pacing, stylet-directed pacing lead
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要