Development of a dosimetry system for therapeutic X-rays using a flexible amorphous silicon thin-film solar cell with a scintillator screen

MEDICAL PHYSICS(2022)

引用 4|浏览17
暂无评分
摘要
Purpose To evaluate the dosimetric characteristics and applications of a dosimetry system composed of a flexible amorphous silicon thin-film solar cell and scintillator screen (STFSC-SS) for therapeutic X-rays. Methods The real-time dosimetry system was composed of a flexible a-Si thin-film solar cell (0.2-mm thick), a scintillator screen to increase its efficiency, and an electrometer to measure the generated charge. The dosimetric characteristics of the developed system were evaluated including its energy dependence, dose linearity, and angular dependence. Calibration factors for the signal measured by the system and absorbed dose-to-water were obtained by setting reference conditions. The application and correction accuracy of the developed system were evaluated by comparing the absorbed dose-to-water measured using a patient treatment beam with that measured using the ion chamber. Results The responses of STFSC-SS to energy, field size, depth, and source-to-surface distance (SSD) were more dependent on measurement conditions than were the responses of the ion chamber, although the former dependence was due to the scintillator screen, not the solar cell. The signals of STFSC-SS were also dependent on dose rate, while the responses of solar cell alone and scintillator screen were not dependent on dose rate. The scintillator screen reduced the output of solar cell at 6 and 15 MV by 0.60 and 0.55%, respectively. The different absorbed dose-to-water measured using STFSC-SS for patient treatment beam differed by 0.4% compared to those measured using the ionization chamber. The uncertainties of the developed system for 6 and 15 MV photon beams were 1.8 and 1.7%, respectively, confirming the accuracy and applicability of this system. Conclusions The thin-film solar cell-based detector developed in this study can accurately measure absorbed dose-to-water. The increased signal resulting from the use of the scintillator screen is advantageous for measuring low doses and stable signal output. In addition, this system is flexible, making it applicable to curved surfaces, such as a patient's body, and is cost-effective.
更多
查看译文
关键词
amorphous silicon thin-film solar cell, dose evaluation, real time dosimetry, scintillator screen
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要