Chromatic drift of the Espresso Fabry-Perot etalon

ASTRONOMY & ASTROPHYSICS(2022)

引用 4|浏览15
暂无评分
摘要
In the last decade, white-light illuminated Fabry-Perot interferometers have been established as a widely used, relatively simple, reliable, and cost-effective way to precisely calibrate high-resolution echelle spectrographs. However, a recent study reported a chromatic drift of the Fabry-Perot interferometer installed at the Habitable-zone Planet Finder spectrograph. In particular, they found that the variation of the etalon effective gap size is not achromatic, as has usually been assumed, but that, in fact, it depends on wavelength. Here, we present a similar study of the Espresso Fabry-Perot interferometer. Using daily calibrations spanning a period of over 2.5 yr, we also find clear evidence for a chromatic drift with an amplitude of a few cm s(-1) per day with a characteristic, quasi-oscillatory dependence on wavelength. We conclude that this effect is probably caused by the aging of the dielectric mirror coatings and we expect that similar chromatic drifts might affect all Fabry-Perot interferometers used for the calibration of astronomical spectrographs. However, we also demonstrate that the chromatic drift can be measured and, in principle, corrected using only standard calibrations based on hollow cathode lamp spectra.
更多
查看译文
关键词
instrumentation, spectrographs, techniques, spectroscopic, techniques, radial velocities
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper