Foley catheter balloon rupture and risk of free fragment formation

JOURNAL OF UROLOGY(2022)

引用 0|浏览9
暂无评分
摘要
Objectives To assess the volume of foley catheter balloon rupture and the incidence of free fragment formation. Materials and Methods Forty Latex and 30 Silicone catheters of common sizes (14-20 French) were inflated with sterile water in an aqueous environment without positive pressure. Catheters producing fragments, fragment size, and maximum instilled volume prior to rupture (“burst volume”) were recorded. A subset of catheters was inflated to recommended balloon volume and the balloon was needle punctured. Results Of the 40 latex catheters, 32 (80%) produced a free fragment during spontaneous rupture. Fragments ranged from 2.1 to 3.2 cm with an average size of 2.74 ± 0.33 cm. Latex catheters had average burst volume of 83 mL, 90 mL, 112 mL,120 mL, and 422 mL for 14 Fr, 16 Fr, 18 Fr, 16 Fr 3-way, and 20 Fr 30 cc, respectively. Of the 30 silicone catheters, spontaneous rupture produced no fragments. Average burst volume for silicone catheters was 57 mL, 45 mL, and 55 mL for 14 Fr, 16 Fr, and 18 Fr, respectively. No catheter balloons produced fragments when needle punctures at recommended balloon volumes. Conclusion Latex catheter balloons tolerate higher fill volumes but have a high likelihood of releasing fragments. Silicone catheters burst at lower volumes, but do not release fragments. At normal fill volumes, needle puncture is safe, and cystoscopy is likely unnecessary.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper